Care & Maintenance Of Instruments
- Clean and lubricate the brand new Hospitime instruments before the first sterilization.
- After use, distinct, clean, lubricate and sterilize the instruments as soon as possible.
- Disinfect chemothermally, thermally or by immersing into a combined disinfecting & cleaning solution.
- Delicate instruments can be cleaned by ultrasonic treatment.
- Rinse thoroughly and carefully after cleaning if possible, use demineralised water.
- Dry sufficiently after rinsing
- Treat the instruments with joints, ratchets or threads with lubricating agent based on paraffin oil.
- We strongly recommend to follow the sterilization instructions of the manufacturer.
- Prior to sterilization, only close the first ratchet on the instruments.
- Normally, autoclaving is performed with saturated water stream at 134ºC.
- When instruments are hot-air sterilized, the temperature should be between 180ºC and 200ºC.
- Corroded instruments should be discarded immediately